Subpixel resolution in CdTe Timepix3 pixel detectors

Research output: Contribution to journalJournal articleResearchpeer-review

Standard

Subpixel resolution in CdTe Timepix3 pixel detectors. / Khalil, Mohamad; Dreier, Erik Schou; Kehres, Jan; Jakubek, Jan; Olsen, Ulrik Lund.

In: Journal of Synchrotron Radiation, Vol. 25, No. 6, 2018, p. 1650-1657.

Research output: Contribution to journalJournal articleResearchpeer-review

Harvard

Khalil, M, Dreier, ES, Kehres, J, Jakubek, J & Olsen, UL 2018, 'Subpixel resolution in CdTe Timepix3 pixel detectors', Journal of Synchrotron Radiation, vol. 25, no. 6, pp. 1650-1657. https://doi.org/10.1107/S1600577518013838

APA

Khalil, M., Dreier, E. S., Kehres, J., Jakubek, J., & Olsen, U. L. (2018). Subpixel resolution in CdTe Timepix3 pixel detectors. Journal of Synchrotron Radiation, 25(6), 1650-1657. https://doi.org/10.1107/S1600577518013838

Vancouver

Khalil M, Dreier ES, Kehres J, Jakubek J, Olsen UL. Subpixel resolution in CdTe Timepix3 pixel detectors. Journal of Synchrotron Radiation. 2018;25(6):1650-1657. https://doi.org/10.1107/S1600577518013838

Author

Khalil, Mohamad ; Dreier, Erik Schou ; Kehres, Jan ; Jakubek, Jan ; Olsen, Ulrik Lund. / Subpixel resolution in CdTe Timepix3 pixel detectors. In: Journal of Synchrotron Radiation. 2018 ; Vol. 25, No. 6. pp. 1650-1657.

Bibtex

@article{86388cc19c3646e8986b6b75b29e5f19,
title = "Subpixel resolution in CdTe Timepix3 pixel detectors",
abstract = "Timepix3 (256 {\~A}— 256 pixels with a pitch of 55 {\^A}µm) is a hybrid-pixel-detector readout chip that implements a data-driven architecture and is capable of simultaneous time-of-arrival (ToA) and energy (ToT: time-over-threshold) measurements. The ToA information allows the unambiguous identification of pixel clusters belonging to the same X-ray interaction, which allows for full one-by-one detection of photons. The weighted mean of the pixel clusters can be used to measure the subpixel position of an X-ray interaction. An experiment was performed at the European Synchrotron Radiation Facility in Grenoble, France, using a 5 {\^A}µm {\~A}— 5 {\^A}µm pencil beam to scan a CdTe-ADVAPIX-Timepix3 pixel (55 {\^A}µm {\~A}— 55 {\^A}µm) at 8 {\~A}— 8 matrix positions with a step size of 5 {\^A}µm. The head-on scan was carried out at four monochromatic energies: 24, 35, 70 and 120 keV. The subpixel position of every single photon in the beam was constructed using the weighted average of the charge spread of single interactions. Then the subpixel position of the total beam was found by calculating the mean position of all photons. This was carried out for all points in the 8 {\~A}— 8 matrix of beam positions within a single pixel. The optimum conditions for the subpixel measurements are presented with regards to the cluster sizes and beam subpixel position, and the improvement of this technique is evaluated (using the charge sharing of each individual photon to achieve subpixel resolution) versus alternative techniques which compare the intensity ratio between pixels. The best result is achieved at 120 keV, where a beam step of 4.4 {\^A}µm {\^A}± 0.86 {\^A}µm was measured.",
keywords = "CdTe X-ray detectors, charge sharing, hybrid pixel spectral detectors, subpixel spatial resolution",
author = "Mohamad Khalil and Dreier, {Erik Schou} and Jan Kehres and Jan Jakubek and Olsen, {Ulrik Lund}",
year = "2018",
doi = "10.1107/S1600577518013838",
language = "English",
volume = "25",
pages = "1650--1657",
journal = "Journal of Synchrotron Radiation",
issn = "0909-0495",
publisher = "Wiley-Blackwell",
number = "6",

}

RIS

TY - JOUR

T1 - Subpixel resolution in CdTe Timepix3 pixel detectors

AU - Khalil, Mohamad

AU - Dreier, Erik Schou

AU - Kehres, Jan

AU - Jakubek, Jan

AU - Olsen, Ulrik Lund

PY - 2018

Y1 - 2018

N2 - Timepix3 (256 × 256 pixels with a pitch of 55 µm) is a hybrid-pixel-detector readout chip that implements a data-driven architecture and is capable of simultaneous time-of-arrival (ToA) and energy (ToT: time-over-threshold) measurements. The ToA information allows the unambiguous identification of pixel clusters belonging to the same X-ray interaction, which allows for full one-by-one detection of photons. The weighted mean of the pixel clusters can be used to measure the subpixel position of an X-ray interaction. An experiment was performed at the European Synchrotron Radiation Facility in Grenoble, France, using a 5 µm × 5 µm pencil beam to scan a CdTe-ADVAPIX-Timepix3 pixel (55 µm × 55 µm) at 8 × 8 matrix positions with a step size of 5 µm. The head-on scan was carried out at four monochromatic energies: 24, 35, 70 and 120 keV. The subpixel position of every single photon in the beam was constructed using the weighted average of the charge spread of single interactions. Then the subpixel position of the total beam was found by calculating the mean position of all photons. This was carried out for all points in the 8 × 8 matrix of beam positions within a single pixel. The optimum conditions for the subpixel measurements are presented with regards to the cluster sizes and beam subpixel position, and the improvement of this technique is evaluated (using the charge sharing of each individual photon to achieve subpixel resolution) versus alternative techniques which compare the intensity ratio between pixels. The best result is achieved at 120 keV, where a beam step of 4.4 µm ± 0.86 µm was measured.

AB - Timepix3 (256 × 256 pixels with a pitch of 55 µm) is a hybrid-pixel-detector readout chip that implements a data-driven architecture and is capable of simultaneous time-of-arrival (ToA) and energy (ToT: time-over-threshold) measurements. The ToA information allows the unambiguous identification of pixel clusters belonging to the same X-ray interaction, which allows for full one-by-one detection of photons. The weighted mean of the pixel clusters can be used to measure the subpixel position of an X-ray interaction. An experiment was performed at the European Synchrotron Radiation Facility in Grenoble, France, using a 5 µm × 5 µm pencil beam to scan a CdTe-ADVAPIX-Timepix3 pixel (55 µm × 55 µm) at 8 × 8 matrix positions with a step size of 5 µm. The head-on scan was carried out at four monochromatic energies: 24, 35, 70 and 120 keV. The subpixel position of every single photon in the beam was constructed using the weighted average of the charge spread of single interactions. Then the subpixel position of the total beam was found by calculating the mean position of all photons. This was carried out for all points in the 8 × 8 matrix of beam positions within a single pixel. The optimum conditions for the subpixel measurements are presented with regards to the cluster sizes and beam subpixel position, and the improvement of this technique is evaluated (using the charge sharing of each individual photon to achieve subpixel resolution) versus alternative techniques which compare the intensity ratio between pixels. The best result is achieved at 120 keV, where a beam step of 4.4 µm ± 0.86 µm was measured.

KW - CdTe X-ray detectors

KW - charge sharing

KW - hybrid pixel spectral detectors

KW - subpixel spatial resolution

U2 - 10.1107/S1600577518013838

DO - 10.1107/S1600577518013838

M3 - Journal article

C2 - 30407174

AN - SCOPUS:85055750958

VL - 25

SP - 1650

EP - 1657

JO - Journal of Synchrotron Radiation

JF - Journal of Synchrotron Radiation

SN - 0909-0495

IS - 6

ER -

ID: 221672129